GLP Compliance Services for Biological Safety Cabinets, Each

4,468.80

Details

IQ Unpacking, Assembly and Installation Verifies that the equipment, manuals, supplies and any other accessories arrived undamaged as specified in the sales order. Verifies equipment is assembled and installed and any initial diagnostics tests have been performed. Any abnormal event(s) observed during assembly and installation have been documented.Utility, Facility and Environment Verifies that the installation site satisfactorily meets manufacturer-specified environmental requirements.OQSystem Component Information Detailed configuration information for each system component have been recorded.OQ Limits: Record a list of manufacturer recommended limits for ensuring that the system is operating as expected.Equipment Operational Tests and Results: Test important equipment functions to verify that the equipment operates as intended by the manufacturer and required by the user. This is a group of selected important equipment parameters for testing according to the intended use of the equipment. - Key Tests: Filter leakage, Air Velocity, Electrical Safety, Air-Flow Smoke Pattern, KI-DISCUS.

Additional Information

SKU 1214571
UOM Each
UNSPSC 56101530
Manufacturer Part Number IOQPCKE89003610